5.2

Starting CIX ASW

i
5.3

Check System

Perform the
workflow checks the system and the accuracy of the calibrations. A particle
standard device, rather than a sample, is scanned and the objects on it are
detected. The system compares the results of this inspection with the
known dimensions of the particle standard device. If discrepancies are
found, a message suggesting optimization measures appears.
Prerequisite
You can find more information about this workflow and using the software in
i
the software's help document.
Click the [Help] button in the software to open the help document for the
current page.
1. Double click the CIX button on the desktop.
 The software starts.
2. Selected a language for the user interface.
3. To do this, click the
[System
software's start page and select the required language from the list.
[Check
System] workflow before working with the system. This
The system has been calibrated.
For more information on the calibration processes, see the software's
help document.
The particle standard device is clean and free of dust.
The particle standard device is in position 2 of the multi-sample holder.
1. Click the
[Check
System] button on the software's start page.
 You are taken through the
21
OLYMPUS Cleanliness Inspector CIX100
5 Switching on the system
Information] > [Languages] button on the
[Check
System] workflow step-by-step.
System Setup Guide