8.9.2.1
Adjusting the friction in the arm
1
2
Turn the clamping knob (1) to loosen the friction in the arm system. After adjusting the
position of the arm the clamping knob should be tightened.
Tighten the nut (2) shown periodically using 16/17 Spanner.
In case of balancing problem contact Siemens Customer Service.
8.10 Removing the top panel , front cover and X-ray source assembly
8.10.1 Removing the top panel
Turn off the Multimobil 2.5,disconnect the mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to transport position. Lock the Multimobil 2.5 by
applying the brake.
Unscrew the two ornamental screws with 2.5 mm allen key.
Grasp the top panel and lift it (as shown in figure).