Drawell DW-TDZ5-WS Operation Manuals - Page 3
Browse online or download pdf Operation Manuals for Laboratory Equipment Drawell DW-TDZ5-WS. Drawell DW-TDZ5-WS 20 pages. Benchtop low-speed centrifuge
Safety Notices
We appreciate your business with us. To prevent any potential accident,
please operate centrifuges according to the following safety protocol.
Safety in installation and maintenance
* The electricity should be cut off before repairing. Please take care if
instrument must be repaired with electricity on. Before opening any lids, please
cut off the electricity and wait for 4 minutes.
* Make sure for placing centrifuge on a steady floor or table and adjusting the
level for normal operation.
* The power supply of centrifuge is the type of nameplate. If you have any
problem of the type of power, please consult the dealer and power supply
bureau in your place.
* As safety measures, the centrifuge equipped with a single-phase
three-phase cable, and the longer cable connects with casing as the protection
grounding. If your sockets hasn't protection grounding, please contact your
electrician to replace the original socket without allowing security role of
single-phase with three-wire power cable to be invalid.
* When replacing parts, the serviceman should use parts manufacturer
required with the same function of original parts. The unrecognized alternative
parts many cause fire, electric shock or other hazards.
* The room should be installed an air-conditioner for good refrigerating effect
when temperature is higher than 35.℃.
* Please keep cleaning of the installation site for the centrifuge. Pay attention
to the blockage of condenser, it should be cleaned once in three months for
guarantee the refrigerating effect.
* Don't unpack and repair instrument by oneself. Please contact with our
technicians for necessary matter.
Attention for fireproofing
* Small breaker (Air switch) can protect the circuit when centrifuge
over-currents. Please use breaker (air switch) with same specification and
1