Auber Instruments SYL-3615 Instruction Manual - Page 4

Browse online or download pdf Instruction Manual for Controller Auber Instruments SYL-3615. Auber Instruments SYL-3615 7 pages. Wi-fi bbq controller
Also for Auber Instruments SYL-3615: Instruction Manual (5 pages)

Auber Instruments SYL-3615 Instruction Manual
AUBER INSTRUMENTS
Start
Step x
Pit
Ending?
Cooking 
Time
Food 
Temp.
Timer
Step (x+1)
End
Figure 8. Flow chart on how the parameters of a cooking step

6. Accessing the Menu

To access the menu, press and hold the SET key for 2 seconds.
6.1 Control Settings
The parameters listed in the Control Setting menu are provided in the table
below. Refer to the following notes for details on each parameter.
Table 3. Parameters in Control Setting.
Name
Description
Pit-AH
Pit High Alarm
Pit-AL
Pit Low Alarm
Food-AH
Food High Alarm
Probe3-AH
Probe 3 High Alarm
Proportional
P
Constant
I
Integral Time
D
Derivative Time
T
Control Cycle Time
MODE
Control Mode
2024.01
Determine the desired pit 
temperature of this step
How do you want to end 
this step?
End the 
program
Enter step 
time or 
Food
target food 
temp.
works.
Range
Default
0 ~ 999
350
0 ~ 999
50
0 ~ 999
350
0 ~ 999
350
1 ~ 999
45
0 ~ 9999
1200
0 ~ 999
75
2 ~ 200
15
Manual-FS,
Manual-TP,
PID-FS
PID-FS,
PID-TP
Save
Save PID Settings
Recall
Recall PID Settings
Back
Note 1. Pit-AH is the high-limit alarm for the pit probe temperature. When the
pit temperature exceeds the Pit-AH value, the alarm is triggered. A 3°
hysteresis band is incorporated for temperature alarms. Users can mute the
device's buzzer by pressing the DOWN (▼) arrow key or using the Mute icon
on the AuberWifi app.
Note 2. Pit-AL is the low-limit alarm for the pit probe temperature. When the
pit temperature falls below the Pit-AH value, the alarm is triggered. A 3°
hysteresis band is applied to temperature alarms. It's important to note that
the Pit-AL alarm is suppressed when the controller is just powered up and has
not yet reached the set temperature.
Note 3. Food-AH and Probe3-AH serve as high-limit temperature alarms for
the Food Probe (Probe 2) and Probe 3, respectively. They operate similarly to
the Pit-AH.
are the high-limit temperature alarm for Food Probe and Probe 3 respectively.
They work in the same way as the Pit-AH.
Note 4. P is the proportional constant, measured in degrees Fahrenheit (°F)
or Celsius (°C). This parameter determines the proportional output based on
the difference between the measured and set temperatures. A larger P value
signifies a weaker action or lower gain. For example, if P = 7, the proportional
band is 7 degrees. When the temperature reading is 10 degrees below the set
value (SV), which is 3 degrees outside the proportional band, the proportional
part of the output will be 100%. If the sensor reading is 5 degrees below the
set value (SV), 2 degrees above the proportional band, the proportional part of
the output will be 71%. When the sensor reading equals the set point, the
controller will have 0% output from its proportional part. This constant also
affects both integral and derivative actions. A smaller P value makes both
Note
integral and derivative actions stronger. It's important to note that the P value
1
is temperature unit-sensitive. If an optimized P value is found when operating
2
the controller in Fahrenheit, one needs to divide the P value by 1.8 when
3
changing the temperature unit to Celsius.
3
4
Note 5. I is the Integral Time, measured in seconds. This parameter
determines the integral output based on the difference between the measured
5
and set temperature integrated over time. For instance, if I = 1000, it means
6
that if the temperature difference between the pit temperature and set
7
temperature remains constant, the output will double after 1000 seconds.
Integral action is utilized to eliminate temperature offset. A larger value for I
8
results in a slower integral action.
WWW.AUBERINS.COM
(Back), BGE,
WSM22,
WSM18,
4,5,6
(Back), BGE,
WSM22,
WSM18,
4,5,6
Back to the main
menu
Back
9
Back
9
P4/7