Olympus UHI-2 Maintenance Manual - Page 27

Browse online or download pdf Maintenance Manual for Medical Equipment Olympus UHI-2. Olympus UHI-2 38 pages.

Olympus UHI-2 Maintenance Manual

3-7 Electric pneumatic unit

Connect to the

manifold unit.

3-8 Manifold unit
CCUK3x6SZ
Connect to the
UPUH15U
pressure sensor.
Connect to the
UPUH15U J10.
DISASSEMBLING/ASSEMBLING PROCEDURE
Connect to the 1st
pressure reducing
unit.
Connect to the
UPUH15U J7.
Connect to the air feed
mouthpiece.
Connect to the UH15U flow
rate sensor.
Connect to the
electric
pneumatic unit.
3-6
Disassembling
1.
Remove the tube and harness from the
electric pneumatic unit.
2.
Remove the screw CCUK4x6SZ (4 pcs.)
which secure the electric pneumatic unit.
Philips screwdriver No. 2
Assembling
1.
Secure the electric pneumatic unit with the
screw CCUK4x6SZ (4 pcs.).
2.
Connect the tube to the electric pneumatic
unit.
Connect the tube correctly according to
the piping diagram.
3.
Connect the harness to the UPUH15U J7.
Philips screwdriver No. 2
Disassembling
1.
Remove the tube from the manifold unit.
2.
Remove the harness from the UPUH15U.
3.
Remove the screw CCUK3x6SZ (4 pcs.)
which secure the manifold unit.
Philips screwdriver No. 2
Assembling
1.
Secure the manifold unit with the screw
CCUK3x6SZ (4 pcs.).
2. Connect the harness to the UPUH15U JlO.
3.
Connect the tube to the manifold unit.
Philips screwdriver No. 2