Olympus UHI-2 Maintenance Manual - Page 28
Browse online or download pdf Maintenance Manual for Medical Equipment Olympus UHI-2. Olympus UHI-2 38 pages.
st
1
pressure reducing
unit
Electric pneumatic
unit
K-connector unit
3-9 Front panel unit
HCBK3x6SA
Spacer 1
DISASSEMBLING/ASSEMBLING PROCEDURE
UHI-2 Piping Diagram
UPUH15U
UPUH25U
Spacer
3-7
UPUH15U flow rate sensor
Downward, Pressure sensor
Manifold unit
Disassembling
1.
Remove the screw HCBK3x6SA (6 pcs.)
which secure the UPUH15U.
2.
Remove the claw of the spacer which
secures the UPUH15U.
3. Remove the spacer 1 (6 pcs.) which secure
the UPUH25U.
Spanner
Philips screwdriver No. 2
Assembling
1. Secure the UPUH25U with the spacer 1 (6
pcs.). Lay the M-sheet in the position
specified on the left when securing the
UPUH25U.
2.
Secure the UPUH15U with the screw
HCBK3x6SA (6 pcs.).
Spanner
Philips screwdriver No. 2