Fig. 2
6. Rotate the upper section of the U-CT30 centering telescope
and bring the bright ring (condenser ring slit) and dark ring
(objective phase plate) into focus.
7. Use the optical element centering knob to center the phase
contrast ring so that the bright ring overlaps the dark ring within
the field of view. (Fig. 2)
# If a multiple number of ring slit images appear, select the
brightest ring to overlap with the phase plate.
8. Repeat steps 6 and 7 for each phase contrast ring.
9. Remove the U-CT30 centering telescope and replace it with
the eyepiece.
10. Widen the field iris diaphragm opening until the diaphragm
image circumscribes the field of view.
}If increased contrast is required, engage the IF550 green inter-
ference filter in the light path.
U-UCD8A