Olympus U-UCD8A Instructions Manual - Page 15
Browse online or download pdf Instructions Manual for Laboratory Equipment Olympus U-UCD8A. Olympus U-UCD8A 40 pages. Motorized universal condenser
Installation using U-DICTS, U-DICTHC or U-DICTHR
Analyzer
(U-AN)
To be inserted into the
mount on the reflected
light illuminator
Analyzer for
transmitted light*
(U-ANT)
To be inserted into the dummy slider
of the U-KPA simple polarized ligh
attachment
* When the U-ANT is used in combination with the U-KPA, the field number becomes 22. If super-widefield
observation (field number 26.5) is required without using a vertical illuminator, use one of the following combi-
nations:
@Attaching the U-AN to the MX-AFDIC analyzer unit.
²Attaching the U-AN360P rotary polarized light analyzer to the U-OPA polarized light attachment.
When an intermediate attachment is used, align the tube clamping knob on the microscope
NOTE
frame with the clamping knob on the attachment in order to determine the orientation of the
analyzer.
** Clamp the slider clamping knobs with a light force. If it is tightened excessively, the prism position switch knob
cannot be operated.
Clamping knob**
Prism position switch knob
Pushed in: BFP1
Pulled out: NORMAL
Transmitted light DIC slider
(U-DICTS, U-DICTHC or U-DICTHR)
Prism control knob
U-UCD8A
DIC-compatible
revolving nosepiece
(U-D6RE, U-D5BDRE, etc.)
12