TEST PROGRAM
©
TEST - OF: Initializing memory (EEPROM)
Press this
key to quit
TEST-OF.
♦ This test initializes the EEPROM that stores the model eye count tables and offset
values by storing their system standard values. This test must always be conducted
whenever the EEPROM is replaced with a new one.
* Accidental execution of this test will overwrite all existing data with standard
values, resulting in accuracy deterioration in any subsequent measurements.
• Initializing EEPROM
Press the measurement button, and the confirmation screen will appear.
[YES] key : Press this key after setting DIP switch No. 5 on the right side of the
main body to ON to start initializing the EEPROM.
When initialization is finished, ''SET DIP SW5 TO OFF" is
displayed.
Return DIP SW No.5 to the OFF position.
[NO] key : Press this key to cancel the initialization test.
The NR - 5500 will then quit TEST - OF.
[6-1 2]