Gainlab Audio DICTATOR GLA-TC1 Kalibratiehandleiding - Pagina 9
Blader online of download pdf Kalibratiehandleiding voor {categorie_naam} Gainlab Audio DICTATOR GLA-TC1. Gainlab Audio DICTATOR GLA-TC1 18 pagina's.
Now go back to 3.1 and check if in Bypass mode you still measure 0V in between TP1 and TP2.
Most likely it will stay as you set up earlier. Repeat 3.1 and 3.2 if needed.
3.3 Bypass measurement (Right channel)
Switch Mode to Bypass.
Measurement process: Measure voltage between TP4 and TP5
points
Adjust: VR4
Target: to see 0V on volt meter
Error margin: 0.1V
3.4 Hicomp measurement (Right channel)
Switch Mode to HiComp.
Measurement process: Measure voltage between TP4 and TP5
Adjust: VR10
Target: to see 0V on volt meter
Error margin: 0.1V