DIEBOLD HSK-100 Gebruikershandleiding - Pagina 13

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} DIEBOLD HSK-100. DIEBOLD HSK-100 14 pagina's. Cam angle gauge
Ook voor DIEBOLD HSK-100: Gebruikershandleiding (16 pagina's)

DIEBOLD HSK-100 Gebruikershandleiding
Spannschulter-Messgerät /
Cam Angle Gauge
HSK 25 - HSK 100
7.

Indicator

7.1.

Disassembling

When the indicator needs to be replaced, remove the adjustment stop screw (sealed
when new) and remove the indicator. Remove the probe and bolt it into the new
indicator.
7.2.

Installation

Install the indicator with the probe installed into the bore and give the indicator a
preload of 2,5 mm. Clamp indicator in this position. Check the function and repeatability
of the indicator.
Attention!
Do not block the indicator by over-tightening.
After replacement, the adjustment of the indicator must be checked with the master
taper (see section Gauge setting)
Spannschulter-Messgerät_D+E
Copyright © Diebold 2012-12 Ersteller: kh
Seite 13 von 14