5.2
Leak testing of containment system
5.2.1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
2, 4
Connections of sample gas path
11, 12
3
Connection of reference gas (if present)
21
BI
Blanking plug
Figure 5-1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
Install the test setup as follows:
1. Close the sample gas outlet
2. If the analyzer has a reference gas connection, also close its inlet
3. Install a suitable shutoff valve (e.g. needle valve V1) of sufficient tightness into the sample gas line upstream of the
sample gas inlet
.
4. Connect a pressure regulator with a measuring range of 200 kPa (OXYMAT 6F) or 50 kPa (CALOMAT 6F) and a
resolution of 0.1 kPa upstream of the shutoff valve V1.
30
using a blanking plug.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
M
Manometer (gauge pressure)
PR
Pressure regulator
V1
Shut-off valve (leak rate < 10
using a blanking plug.
kPa*l/s)
-6