Siemens ULTRAMAT 6F Manual de Instruções - Página 30

Procurar online ou descarregar pdf Manual de Instruções para Equipamento industrial Siemens ULTRAMAT 6F. Siemens ULTRAMAT 6F 49 páginas. Gas analyzer for the determination of ir-absorbing gases
Também para Siemens ULTRAMAT 6F: Manual de equipamento adicional (20 páginas), Manual de Segurança (24 páginas)

Siemens ULTRAMAT 6F Manual de Instruções
5.2
Leak testing of containment system
5.2.1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
2, 4
Connections of sample gas path
11, 12
3
Connection of reference gas (if present)
21
BI
Blanking plug
Figure 5-1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
Install the test setup as follows:
1. Close the sample gas outlet
2. If the analyzer has a reference gas connection, also close its inlet
3. Install a suitable shutoff valve (e.g. needle valve V1) of sufficient tightness into the sample gas line upstream of the
sample gas inlet
.
4. Connect a pressure regulator with a measuring range of 200 kPa (OXYMAT 6F) or 50 kPa (CALOMAT 6F) and a
resolution of 0.1 kPa upstream of the shutoff valve V1.
30
using a blanking plug.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
M
Manometer (gauge pressure)
PR
Pressure regulator
V1
Shut-off valve (leak rate < 10
using a blanking plug.
kPa*l/s)
-6
A5E45779389002-01, 09/2018