Siemens ULTRAMAT 6F 사용 설명서 - 페이지 30

{카테고리_이름} Siemens ULTRAMAT 6F에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Siemens ULTRAMAT 6F 49 페이지. Gas analyzer for the determination of ir-absorbing gases
Siemens ULTRAMAT 6F에 대해서도 마찬가지입니다: 추가 장비 매뉴얼 (20 페이지), 안전 매뉴얼 (24 페이지)

Siemens ULTRAMAT 6F 사용 설명서
5.2
Leak testing of containment system
5.2.1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
2, 4
Connections of sample gas path
11, 12
3
Connection of reference gas (if present)
21
BI
Blanking plug
Figure 5-1
Recommended test setup for CALOMAT 6F, CALOMAT 62F and OXYMAT 6F
Install the test setup as follows:
1. Close the sample gas outlet
2. If the analyzer has a reference gas connection, also close its inlet
3. Install a suitable shutoff valve (e.g. needle valve V1) of sufficient tightness into the sample gas line upstream of the
sample gas inlet
.
4. Connect a pressure regulator with a measuring range of 200 kPa (OXYMAT 6F) or 50 kPa (CALOMAT 6F) and a
resolution of 0.1 kPa upstream of the shutoff valve V1.
30
using a blanking plug.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
M
Manometer (gauge pressure)
PR
Pressure regulator
V1
Shut-off valve (leak rate < 10
using a blanking plug.
kPa*l/s)
-6
A5E45779389002-01, 09/2018