Section 3. Installation

Now, it's time to connect the SNAP Vacuum Controller: The only required connections
for operation are power, a vacuum pump, and a vacuum system.
Hooking the SNAP up to a system is a simple process:
1.Mounting: Find a suitable location for
mounting the SNAP. In its standard,
landscape mode, SNAP sits on rubber feet.
It should be put in a place that easily
allows access to the touch screen, but has
minimal process vibration. A mounting
bracket or pole clamp can be added-on to
allow for direct mounting to certain
vacuum ovens or rotovaps. These options
are best used in SNAP's portrait mode.
2.Connect to Vacuum Pump: Connect your
vacuum pump to the KF25 input labeled
"Vacuum Pump".
3.Connect to Vacuum System: Connect
your vacuum system to the KF25 vacuum
output labeled "System".
4.Vent Port (optional): If you would like to use a gas other than ambient air to vent
your system, connect that vent gas to the hose barb.. Many customers may
choose to vent with dry Nitrogen.
5.Data Capture (optional): Attach USB connection for data to a PC or other system.
6.Analog In/Out (optional): Connect to a PLC and use specified voltages to control.
7.External Sensor (optional): Connect an external sensor(e.g. 10 Torr Capacitance
Manometer for improved accuracy
8.Power: Plug in to Power and turn on
©
2021 DigiVac Company®Allrights reserved. | 1020 Campus Drive, Morganville,NJ 07751 | (732) 765-0900