ND filter sliders
Epi-illumination attachment
Brightness
1
1/4
1/8
1/16
1/32
1/64
1/128
1/512

Transmitted image in fluorescence observation

2.5
For fluorescence observations, turn off the microscope power switch to cancel the transmitted
image.
Bright ambient lights will make it more difficult to view the image. We recommend keeping the
room dark during fluorescence observations.

Adjusting the Epi-fluorescence Aperture Diaphragm

2.6
In Epi-fluorescence microscopy, image brightness can be adjusted by varying aperture
diaphragm size. Stopping down the aperture diaphragm darkens the image, while opening up
the aperture diaphragm makes the image brighter.
Use the aperture diaphragm of the Epi-fluorescence attachment or DIH-M after performing the
centering procedures. For a detailed discussion of this topic, refer to "17.5 Epi-fluorescence
Aperture Diaphragm."
Camera adjustment
2.7
(adjusting the brightness of the image on the monitor)
When observing images captured by the camera and displayed on the monitor, you can adjust
brightness by varying camera adjustment parameters, such as display mode, exposure mode,
metering mode, exposure compensation, and image level adjustment.
For detailed discussion, refer to the operating manual provided with the camera or camera
control software.
Individual Operations
2 Brightness Adjustment
ND4
ND8
-
-
-
-
-
-
-
-
12
ND filter sliders
DIH-M
ND16
-
-
-
-