Drucker Diagnostics 614M Instrukcja obsługi - Strona 2
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt laboratoryjny Drucker Diagnostics 614M. Drucker Diagnostics 614M 12 stron. 12vdc mobile labaratory centrifuge
Table of Contents
Model Description ............................... pg. 2
Intended Use ....................................... pg. 2
Supplied Equipment ............................ pg. 2
Technical Specifications ...................... pg. 3
Initial Setup Procedure ........................ pg. 3
Front Panel Controls ........................... pg. 4
Operation ............................................ pg. 5
Tube Holder Configurations ................ pg. 6
Rotor Removal and Installation ........... pg. 7
Care and Maintenance ........................ pg. 7
Cleaning and Disinfection ................... pg. 8
Troubleshooting ................................... pg. 9
Safety .................................................. pg. 10
Emergency Rotor Chamber Entry ....... pg. 10
Calibration and Ground Testing ........... pg. 11
Replacement Parts .............................. pg. 12
Available Accessories ......................... pg. 12
Model Description:
The Model 614M is a 12VDC continuous-duty, electronically-controlled fixed-angle
centrifuge intended for mobile applications where the Vehicle has a 12VDC power
supply (cigarette lighter). The unit is controlled by a mechanical timer settable from 1
to 30 minutes. Samples can be safely viewed through the transparent lid. In the event
that the lid is opened during a run, the power to the motor is disconnected.
For warranty information, turn to page 12.
Intended Use:
This centrifuge is an IVD accessory, and therefore subject to the former EU IVD
Directive 98/79/EC and the current EU IVD Regulation 2017/746. It is a laboratory
product designed to separate components by generation of relative centrifugal force.
It separates human and animal samples - such as blood, urine and other body fluids
-- collected in appropriate specimen receptacles, either alone or with reagents or other
additives. As a general purpose laboratory centrifuge, it is designed to also run other
containers filled with chemicals (non-flammable, non-explosive, non-volatile, and non-
highly reactive only), environmental samples, and other non-human body samples.
This centrifuge should be operated by trained personnel only. Any use other than
those intended by the Manufacturer is explicitly prohibited. [Maximum sample density
is 1.15 grams/mL].
Supplied Equipment:
The following items come standard with each Model 614M centrifuge:
The rotor and rotor accessories are rated for a rotation frequency of 4,000 RPM.
1.
One (1) six-place fixed-angle rotor
2.
Six (6) 125 mm tube holders
3.
Six (6) 75/100 mm tube holders
WARNING: Use universal precautions when
handling laboratory specimens. All human specimens
of blood, blood compounds, or bo-dily fluids are
potentially biohazardous and may contain hepatitis
B virus, hepatitis C virus, human immunodeficiency
virus or other disease causing agents. For the safety of
both the operator and service personnel, care should
be taken when handling substances that are known to
be toxic, radioactive, or contaminated with pathogenic
microorganisms when using this centrifuge. When
Risk Group II materials are used
(As identified in the World Health Organization
"Laboratory Bio-Safety Manual") the use of a Bio-Seal
should be employed. More than one level of protection
must be provided in the case of materials of a higher
group. The use of flammable or explosive materials
as well as those materials which chemically react
vigorously is prohibited. Any use of the equipment in a
manner not specified in these instructions may impair
the level protection provided by the equipment.
p/n
7786047
p/n
7713032
p/n
7713079
Page 2