Use of an eyepiece micrometer
I
An eyepiece micrometer (10
mm/100~
(optionally available) can be
inserted into the eyepiece CWHK10X in the following procedure:
1) Remove the retaining ring'
1./
from the lower end of the eyepiece
and place the micrometer on the retaining ring with the reticle-
engraved surface, facing downward. (Fig. 2)
*
Be certain to clean the micrometer disc before inserting into the
eyepiece.
2)
Return the retaining ring into the eyepiece and insert the eyepiece
into the eyepiece tube.
Fig.
3
Fig.
4
Fig. 5
III
Mounting the objectives
1) Lower the stage by means of the coarse adjustment knobs.
2) Screw the objectives into the nosepiece, from low power to higher
power in a clockwise direction. (Fig.
3)
*
Close the empty aperture in the nosepiece with a plug
CD
provided.
(Fig. 3)
II
Mounting the mechanical stage
Mount the mechanical stage on the plain stage
CD
in a manner that
the specimen traversing guide of the mechanical stage is located
closest to the pillar, and tighten the stage clamping screws
®
with
a coin. (Fig. 4)
• Insertion of stage clips
The plain stage is pre-drilled for insertion of the stage clips when the
mechanical stage is not attached (see page 4).
II
Mounting the condenser
Insert the condenser into the condenser holder, with the condenser
iris diaphragm lever
(1),
directing in the microscope front, and
tighten the clamping screw (2) . (Fig. 5)