Ega Master 57666 Instrukcja obsługi - Strona 10

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Przyrządy pomiarowe Ega Master 57666. Ega Master 57666 18 stron. Máquina de calibración de par

Ega Master 57666 Instrukcja obsługi
1
.
Turn the meter on and select the desired torque.
2
.
Place the torque wrench in the housing.
3
.
We turn the crank to the left until the torque shown by the meter drops to the
desired torque (if you want to raise the torque, turn the crank to the right).
1
Remark: For making correct calibration test, it is advised to follow international
standard procedures: for example, international standard ISO6789 or similar.
Important: Overload in the toque tester can damage it.
Keep the machine clean.
Perform periodic torque tester calibration.
10
QUICK USE GUIDE
Before starting the calibration:
-
the torque wrench must not be loaded.
-
make sure the torque tester is on when calibrating.
2
MAINTENANCE
3