ANDalyze AND1100 Manual do utilizador - Página 5

Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição ANDalyze AND1100. ANDalyze AND1100 16 páginas. Fluorimeter for water testing

AND1100 Fluorimeter for Water Testing
 
3.2  General Features 
 
Fluorescence Based Sensing ‐ Reaction produces quantitative fluorescence based results. 
Sensor Kit ‐ Each disposable color‐coded sensor is designed for a specific heavy metal test target. Kits 
provide sampling tubes, syringes and cuvettes. 
Rugged & Convenient Construction ‐ Tough, light weight, impact‐resistant shell protects electronics and 
sensors. 
Key Pad ‐ Quick, simple menu navigation and one button push for sample analysis. 
Data Capture & Reporting ‐ Time and date stamped with sample number. Stores over 240,000 test results 
and over 75 site locations.  Data downloadable through a USB cable. 
Water and Dust Resistant  ‐ Enclosure rating ‐ IP54 
Portable ‐ Battery operated and rechargeable through USB cable or traditional adapter. 
 
Note: See full technical specifications in section 9. 
 
3.3  Analyte Sensors 
 
To help differentiate the materials specific to each analyte, the packaging has been individually labeled and the 
sensor housings and buffer caps have been color coded.  For Example: 
 
Green – Lead (Pb)     Orange – Uranium (U)       Blue – Copper (Cu)      Gray – Mercury (Hg)    Etc. 
 
Note: In some cases the same sensor is used for multiple ranges but different buffers and/or methods will 
accompany the sensor in the ki
 
3.4  Operating Environment and Storage 
 
Important:  The  following  conditions  are  necessary  to  ensure  correct  instrument  operation  and  accurate 
results: 
 
Place the instrument firmly on a flat and even surface during operation. 
Maintain an ambient temperature of 10 to 40 ºC (50 to 104ºF) for proper fluorimeter operation. Please see 
the manual for each specific metal for sensor specific temperature conditions. 
The relative humidity should be less than 80%; moisture should not condense on the instrument. 
Do not operate or store the instrument in extremely dusty, damp or wet locations. 
Keep the surface of the instrument, the cell compartment and all accessories clean and dry. 
Sensors  are  stable  up  to  1  year  if  stored  at  <23˚C  (74˚F),  <50  %  R.H.  away  from  direct  sunlight.  For  best 
results, store in the refrigerator (4˚C/39˚F). 
Buffer  solutions  are  stable  up  to  6  months  if  stored  at  <23  ˚C  (74˚F)  but  if  stored  in  the  refrigerator 
(4˚C/39˚F), can be used for up to 1 year. 
Sensors and buffers must be brought to room temperature before use. 
ANDalyze Inc. 800 Boylston Street, Prudential Tower, Floor 16, Boston, MA 02199 USA Tel. +1 857.453.6740 www.andalyze.com 
t. 
 
 
Page 5 of 16 
 