AND1100 Fluorimeter for Water Testing
3.5 Water Testing Guidelines
Note: This is a drinking water test kit. If you are testing other matrices such as surface water, ground water or
industrial water; contact ANDalyze customer service for additional application notes.
Water Sampling
‐
For best results use freshly collected sample (unpreserved) for analysis. We recommend that you use the
sample within 1 hour (maximum of 2 hours) of collection to minimize any metal loss to the walls of the
sample container. This is particularly important for testing trace lead levels.
‐
Once the sample is mixed with ANDalyze sample buffer, test within 15 minutes.
pH Range
‐
Tests have shown that environmental samples preserved in acid to a pH < 2 cannot usually be brought to
an appropriate pH when mixed with the ANDalyze buffer. These samples must be first neutralized with
NaOH to a suitable pH before mixing with ANDalyze buffer. Please contact ANDalyze customer service for
instructions related to pre‐treatment of highly acidic or highly basic samples.
4 Set‐Up and Operating Procedures
4.1 Charging the Battery
Important: Before using the instrument assure that the fluorimeter device is charged.
To Confirm the Battery Level:
1. Turn Power ON by pressing and Holding the ON/OFF button. (See
Horizontal Arrow on Photo to Right)
2. Once device is initiated, confirm the battery indicator status is green
(See Vertical Arrow on Photo to Right). The battery indicator should not
be red (as seen in the photo to right)
Charging the Battery:
1. Release the Rubber Boot at the bottom of the meter.
2. Attach the USB to MINI‐B Cable to the outlet at the bottom of the
device and to a USB powered source such as a computer or a USB
charger. (See Photo to Right).
Note: When the meter is charging, the battery indicator will display yellow
lines across and around the indicator.