ADE EHR1 Manual de instruções - Página 11
Procurar online ou descarregar pdf Manual de instruções para Balanças ADE EHR1. ADE EHR1 20 páginas. Platform scale
Automatic Switch-off
For energy saving the scale can be set for automatic switch-off. Automatic switch-off can be
preselected after 1 up to 15 minutes or disabled.
When scale is on, press both button ON/OFF/
appears in display.
Then display shows P1.xx and the first digit is flashing.
By pressing button „T" you can change the digit. The value of the digit can be changed by
pressing button „U".
Both digits together means the remaining time until automatic switch-off.
By preselection of P1.00 automatic switch-off is disabled.
After setting the preselection press button „T" and P2.xx appears in display for preselecting
hold function.
To save the preselection and to leave SETUP-menu press button ON/OFF/
Setting of Hold Function
The scale has a hold function which can be preselected for automatic hold or which van be
disabled.
When scale is on, press both button ON/OFF/
appears in display.
Then display shows P1.xx and the first digit is flashing.
Now press button „T" again, until P2.xx appears in display and the first digit is flashing.
By pressing button „T" you can change the digit. The value of the digit can be changed by
pressing button „U".
By preselection of P2.00 Auto-Hold-Function is enabled.
Further preselections are P2.01 (automatic hold, until larger weight reading), P2.02 (automatic
hold, when weight is below 10d and auto-hold new stable weight is more than 10d), P2.03...50
(automatic hold, when unchangeable reading when the variety is within +/-3d bis 50d).
After setting the preselection press button „T" and P3.xx appears in display for preselecting
RS232 interface settings.
To save the preselection and to leave SETUP-menu press button ON/OFF/
EHR_EHQ-210517-Rev009-UM-en
and
„U" same time as long as SETUP
and
„U" same time as long as SETUP
.
.
10