4.2 Selecting the Location

Put the device on a horizontal, leveled, and stable surface.

Leave enough space around the device for normal air circulation (4 in./10 cm minimum).

Leave enough space around the device for easy access and maintenance.

Do not use the device in surroundings where there are major temperature and humidity

fluctuations. Also avoid locations in direct sunlight or places near devices that produce heat.
Avoid locations with excessive vibrations.


4.3 Attaching the Power Cord

Fit the main power cord, which is included in the package, into the power receptacle on the
vortexer. Connect the other end of the cord to a properly grounded wall outlet. To avoid
interference from noise, surges, and spikes, a dedicated line is preferred. If no such line is
available, avoid lines to which powerful electric motors, refrigerators, and similar devices are
connected.
Be careful when you plug the cord to a grounded wall socket. Do not touch plug
with wet hands. Do not pull the plug by the cord.

4.4 Installing Vortexer Platforms

Cat. No.
Description
48010
Platform for 4 microplates
48011
Platform with non-slip rubber mat
48012
Spring loaded platform for flasks, bottles, or tube racks
The Corning® LSE Digital Vortexer requires the installation of the selected platform. Please
follow the instructions below for installing the platform.
Before you attach the platform, make sure that the slide stabilizers are properly fastened,
adjusted, and lubricated with the included grease.
Sliding Stabilizers
Grease
Instruction Manual
3