SENC 150 Reading Head Replacement Instructions
Installing alignment brackets
Attach the alignment brackets provided by inserting the top
ears into the bottom grooves in the scale case.
Numbers and letters on reading head casting and alignment
brackets correspond when assembled.
Slide brackets against
reading head snap into place.
Installing end cap(s)
spring clip
Insert brackets at 45°
and twist to lock in scale grove.
Determining cable direction
Determine cabel exit direction before mounting the encoder.
To change the cable exit direction, remove the cover plate
and rotate it 180°.
K1 - cable exit right side
1
1
K2 - cable exit left side
1
1
7 Preparing the remounting
Do not pull out the PCB!
1
T-10 torx screw
Do not clamp the wires!
1
T-10 torx screw
max. 70 Ncm!