X. DC Current (DCA)
1. Press the "
" button or "
nto the terminal of "mA" or "10A". See the picture below.
2. Circularly press the "
automatic DC 600aA and 10A current to test. The "DC" 、 "AUTO" 、
"APO"symbols will appear on the LCD screen.
3.In case "OL" is displayed on the display, it indicates the current being
measured has exceed the current range, and please select higher ranges
for measurements.
Caution:
a)At the 10A mode, current bigger than 10A cannot be measured, and at mA mode, the current bigger than 600 mA cannot
be measured. Otherwise this will lead to the burning of the fuse or damage the instrument.
b) When the meter pen is plugged in the input terminal of the current, it is strictly prohibited to have the meter pen connected
in parallel on any circuits.
Range
60mA
600mA
6A
10A
The maximum input current: 10A (not exceeding 15s))
Overload protection: 0.6A/250V fuse; 10A/250V fuse.
XI. AC Current (ACA)
1. Press the "
" button or "
into the hole of "mA" or "10A". See the picture below.
2. Circularly press the "
"AUTO" 、 "APO" symbols will appear on the LCD screen.
3. In case "OL" is displayed on the display, it indicates the current being measured has exceed the current range, and please
select higher ranges for measurements.
Caution:
a)At the 10A mode, current bigger than 10A cannot be measured, and at mA mode, the current bigger than 600 mA cannot
be measured. Otherwise this will lead to the burning of the fuse or damage the instrument.
b) When the meter penis plugged in the input terminal of the current, it is strictly prohibited to have the meter pen connected
in parallel on any circuits.
Range
60mA
600mA
6A
10A
The maximum input current: 10A (not exceeding 15s))
www.tmatlantic.com
" button. Plug the black meter pen into the terminal of "COM" and the red meter pen
" button or "
Accuracy
Resolution
10μA
±(1.0%+10d)
100μA
1mA
±(1.2%+10d)
10mA
" button. Plug the black meter pen into the hole of "COM" and the red meter pen
" button or "
"button. Select automatic AC 600aA and 10A current to test. The "AC" 、
Accuracy
Resolution
10μA
±(1.5%+10d)
100μA
1mA
±(2.5%+15d)
10mA
"button. Select
- 6 -
10A
switch
400mA
switch