Aktakom ATK-2001 Kullanıcı Kılavuzu - Sayfa 6

Ölçüm Cihazları Aktakom ATK-2001 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Aktakom ATK-2001 11 sayfaları. Clamp meter

Aktakom ATK-2001 Kullanıcı Kılavuzu

3. Operation Instructions

3.1. DC/AC Current Measurements

WARNING: Make sure that all the test leads are disconnected from the meter's
terminals for current measurement.
3.1.1. DC Current
a. Set the rotary switch at mA DC or 40A DC.
b. Push the zero button to adjust the reading to zero. Due to the high
sensitivity of the clamp meter, must zero in the same direction as in
measurement to avoid interference by external magnetic field.
c. Press the trigger to open the jaw and fully enclose the conductor to be
measured. No air gap is allowed between the two half jaws.
d. Read the measured value from the LCD display.
3.1.2. AC Current
a. Set the rotary switch at 400mA AC, 4000mA or 30A AC.
b. Press the trigger to open the jaw and fully enclose the conductor to be
measured. No air gap is allowed between the two half jaws.
c. Read the measured value from the LCD display.
4
www.tmatlantic.com