Examples
Simple setup (and minimal instrument test)
Connect the output of the Model 2100 to an oscilloscope. Before connecting the Model
2100 to any other instrument, or the experimental preparation, it is safest to turn its
power off.
Set the Output controls:
Range:
Pulse Sign:
Baseline:
Pulse Amplitude:
Setting The Timing Controls:
Delay:
Burst Width:
Duration:
Period:
Setting The Start Controls:
Mode:
Disconnect any cables to the TRIG or GATE inputs.
Note that the green Trig and Event indicators turn on as Free Run is turned on (and
turn off otherwise). You should observe a 15ms duration, 20Hz rate pulse. of 2.00V
above a 0V baseline on the oscilloscope.
A-M Systems
Operating Instructions
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
10V
+
Zero
2.00
none
single
-2
1.50 * 10
5.00 * 10
-2
Free Run
http://www.a-msystems.com
6