- ページ 9

モビリティ・エイド Etac Molift Air SeriesのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Etac Molift Air Series 15 ページ。

Etac Molift Air Series ユーザーマニュアル

Mounting Suspension

1. Align suspension in connection point.
2. Push button on locking pin and insert all the
way through.
3. Make sure locking pin is properly fastened
4. Push button and push bolt from opposite side
while holding in button to remove locking pin.

Emergency Stop / Emergency lowering

Emergency Stop
Activation and reset of emergency stop:
Pull to activate emergency stop. The button will
come out, and hoist will stop.
Push with finger or use tube on cord to push but-
ton back in to reset emergency stop.
Electrical Emergency Lowering
Pull and hold to start lowering. Hold until user is
lowered and can be released from suspension.
Only for emergency use when hand
control does not work! Stop before
full band is out, or lifting band will
be winded the wrong way and go
upwards.
Manual Emergency lowering
Allen key or bit for power drill can be found inside
the lifter. Remove cover to locate.
Push down and press cover outwards with a flat
screwdriver in the two slots to open cover. Turn
the "Brake Release" handle down.
Locate sticker over the hole for manual emergency
lowering. Punch hole or remove sticker and insert
tool. Rotate counterclockwise to lower lifting belt.
Using a power drill is recommended.
Lifter must be sent to service after performing
manual emergency lowering.
8