4. zero: ZERO INDICATOR turns on when scale is in Zero position
5. tare: TARE INDICATOR turns on when this function is used
6. LoUW: turns on when sample unit weight is not heavy enough for accurate counting
7. LoWT: turns on when sample number is not enough for accurate counting
8. M+: is used when accumulating the counts data; this data is memorized by pressing this key and
can be accumulated up to 99 times.
9. S1: SCALE 1 working state
10. S2: SCALE 2 working state
11. S3: SCALE 3 working state
12. LoB: low voltage, please charge battery
13. CHG: charged state, red LED lights up when charging, when the battery is fully charged, Red
and Green LED lights flash alternately.

4.1.2 Key functions

1. SCALE/SELECT: Press this button to select available counting scale among scale 1, scale 2, and
scale 3. You can only choose to use one of them at a time.
 SCALE 1: Main counting scale, the bigger platform
 SCALE 2: Smaller counting scale offers more accuracy
 SCALE 3: Extra connection platform. When we get the sample weight from SCALE 2,
we can use this platform to count more samples.
Operation way:press< Scale/ select > key,WEIGHT line will show flashed [SCALE 1],
press
< Scale/ select > key to select then press <Enter> key to confirm .
2. UNIT key changes the weight units between kg and lb
3. ZERO key sets display to zero or subtract the container weight. The zero range is <10% of full
capacity. ZERO LED indicator will illuminate when you active the ZERO function.
4. TARE key subtracts the weight of a container placed on the weighing platform. The tare range is
<100% full capacity. Press the TARE key to enable the Auto-Tare function, the TARE LED
indicator will illuminate when you activate the TARE function.
Operation way 1 : turn on the scale and wait for ZERO mode then place container
press<TARE>key ,WEIGHT line will show flashed 0.0000,TARE LED indicator will light up
when you activate it.
Operation way 2: (new version function)turn on the scale and wait for ZERO mode then
place container press<TARE>key, WEIGHT line will show -X.XXX,ZERO LED indicator will
light up when you activate it.
5. PRINTER key is used to transmit data to a printer, computer, or other device.
It should be connect to PC or printer by USB or RS232.
6. ENTER key has the same function as it would on a computer keyboard.