Autonics krn100 series Manual - Page 8
Browse online or download pdf Manual for Measuring Instruments Autonics krn100 series. Autonics krn100 series 17 pages. 100 mm hybrid recorder
Also for Autonics krn100 series: Instruction Manual (2 pages), Instruction Manual (2 pages), Product Manual (3 pages)
Inside
①
②
Paper cassette
①
⑤
②
Functions
Special function [ Special Function ]
It•displays•the•applied•measuring•value•of•the•set•special•
function.•Depending•on•Input•Type(Input•specification),•
applied•special•function•is•different.
• • • S et•range•:
• -•When•input•type(input•specification)•is•temperature•
sensor•(thermocouple,•RTD):•None•↔•Difference
• -When•input•type(input•specification)•is•analog•(voltage,•
current):•Linear•↔•Root•↔•Square•↔•Two•Unit
• (Two•Unit•is•displayed•when•Input•Type•(input•
specification)•is•set•as•0-20mA,•4-20mA.)
• • Factory•default:•None
٧ Difference (deviation)
It•is•available•to•set•when•Input•Type(input•
specification)•is•temperature•sensor•(thermocouple,•
RTD).•It•displays•the•deviation•of•Reference•Channel•
(Reference•channel)•measuring•value.•
(Display•value•=•standard•channel•measuring•value•-•
reference•channel•measuring•value)
• • The•set•channel•as•analog•(current,•voltage)•of•
Input•Type•(Input•specification)•is•not•able•to•set•as•
Reference•Channel•(reference•channel).
• • • I f•there•is•no•set•reference•channel,•it•displays•
standard•channel•measuring•value.
• • If•any•one•of•reference•channel,•or•standard•channel•
is•break•(BURN),•upper•limit•value•(HHHH),•lower•
limit•value(LLLL)•status,•it•displays•as•correspond•
value.•If•you•select•the•channel•which•is•used•
Difference•function•as•reference•channel,•it•displays•
the•value•based•on•calculating•actual•measuring•
value,•not•display•value•of•reference•channel.
③
④
③
100mm Hybrid Recorder
①•Ink•cartridge(model:•D33006B-66X-01)
②•Recording•paper•cassette
Cassette•saves•the•recording•paper.
③•Recording•paper•cassette•lever
Press•the•lever•down•and•this•recording•paper•cassette•is•
removed•from•KRN100.
※ • R emove•the•recording•paper•cassette•for•recording•paper•
replacement,•ink•cartridge•replacement.
①•Recording•paper•holder
Movement•holder•of•recording•paper•when•recording
② Recording•paper•storage•part
Storage•part•for•recorded•recording•paper
③•Front•cover•of•recording•paper•storage
Open•recording•paper•guide•for•recording•paper•replacement•
④•New•recording•paper•storage:•Storage•part•for•new•
recording•paper•(1EA•recording•paper•is•storable.)
⑤ Rear•cover•of•recording•paper•storage•
٧ Linear
It•applies•lower•limit•scale•and•upper•limit•scale•to•lower•
limit•input•value•and•upper•limit•input•value•and•displays•
this•values.•
Ex)•In•case•low•limit•input•value:•-5V,•high•limit•input•
value:•+5V•and•in•case•lower•limit•scale:•-1000,•
upper•limit•scale:•1000,•if•current•input•value•is•2V,•
display•value•is•400.
٧ Root
In•case•voltage,•current•input•type,•this•mode•is•used•
when•input•value•is•calculated•by•Root•(
desired•display•value.•Differential•pressure•signal•of•
differential•pressure•flow•meter•is•calculated•Root•(
for•the•to-be•measured•flux.•This•function•is•used•to•
measure•flux•by•input•value.
Ex)•In•case•lower•limit•input•value:•-5V,•upper•limit•input•
value:•+5V•and•in•case•lower•limit•scale:•-1000,•
upper•limit•scale:•1000,•if•current•input•value•is•2V,•
display•value•is•approx.•673.32.
٧ Square
In•case•of•voltage,•current•input•type,•this•mode•is•
used•when•input•value•is•calculated•by•square•for•the•
desired•display•value.•Reverse•of•Root,•flux•signal•is•
calculated•by•square•for•differential•pressure•signal.
Ex)•In•case•lower•limit•range:•-5V,•upper•limit•range:•
+5V•and•in•case•lower•limit•scale:•-1000,•upper•
limit•scale:•1000,•if•current•input•value•is•2V,•display•
value•is•-20.
A. Recorder
B.•Indicator
C.•Converter
D.•Controller
E.•Thyristor•
•••••unit
F.•Pressure•
••••transmitter
G.•Temp.•
•••••transmitter
H.•Accessories
KRN100
KRN50
)•for•the•
)
A-9