CORNING LSE 6705 Instruction Manual - Page 5

Browse online or download pdf Instruction Manual for Laboratory Equipment CORNING LSE 6705. CORNING LSE 6705 8 pages. Platform rockers
Also for CORNING LSE 6705: Instruction Manual (12 pages)

CORNING LSE 6705 Instruction Manual

3.3 Adjusting the Angle

Unload the platform. Locate the knob on the right hand side of the rocker. Note the tilt degree
markings on the platform drive rod (front center under platform). Loosen knob on the side by
rotating counterclockwise. Adjust the platform to desired tilt angle. Retighten the knob on
the side by turning clockwise. Test the unit without load to ensure platform is tight.
NOTE: You many need to turn the rocker on and off to change drive rod position before adjusting.

3.4 Attaching the Power Cord

Connect the main power cord to the rocker and then connect the plug end of the cord to a
grounded wall socket. Avoid lines to which powerful electric motors, refrigerators, and similar
devices are connected.
Be careful when you plug the cord to a grounded wall socket. Do not touch plug
with wet hands. Do not pull the plug by the cord.

3.5 Installing Double Stacking Platform

1. Set the posts (A) through the 4 holes at the corners of the bottom platform.
2. Add a washer (B) and nut (C) to each post.
3. Fit the upper platform (D) on the top of the 4 posts through the corner holes.
4. Screw on the 4 metal knurl nuts (E) to secure the upper platform.
NOTE: Misuse or unauthorized modifications will waive the warranty rights. If the equipment
is used in a manner not specified by the manufacturer, the safety features of the rocker may
be impaired.
Instruction Manual
3