Figure 56 – Thoroughly clean slide-mount base (old style)
The older style nosepiece slide mount is now ready to
be reinstalled onto the stand and put back into service
(see
Figure
57).
Figure 57 – Slide-mount parts ready for use (old style)

Servicing the New-Style Slide Mount (BHS/BHT)

This section applies only to the newer style assembly
(see
Figure
49). The older style assembly is covered in
the previous section.

Disassemble the New-Style Slide Mount (BHS/BHT)

Using a suitable slotted screwdriver, loosen and remove
the stop screw for the nosepiece slide-mount locking
block (see
Figure
58).
Figure 58 – Remove the locking-block stop screw
Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames
Using a suitable slotted screwdriver, loosen and remove
the pivot screw for the nosepiece slide-mount locking
block (see
Figure
59).
Figure 59 – Remove the locking-block pivot screw
Separate the locking block from the slide-mount base
(see
Figure
60).
Figure 60 – Separate locking block from slide-mount base

Clean Components of New-Style Slide Mount (BHS/BHT)

Using a cotton swab and a suitable solvent (acetone),
thoroughly clean the nosepiece slide-mount base,
paying special attention to the surface upon which the
locking block slides and the dovetail bearing surfaces.
(see
Figure
61).
Figure 61 – Thoroughly clean slide-mount base (new style)
Revision 1