Olympus IX70 Panduan Pengoperasian - Halaman 31

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Mikroskop Olympus IX70. Olympus IX70 32 halaman. Inverted fluorescent microscope
Juga untuk Olympus IX70: Petunjuk Pengoperasian (5 halaman)

Olympus IX70 Panduan Pengoperasian

Filter cube specs:

DAPI (quad)
DAPI (oly MNU2)
CFP
FITC (wide)
FITC(narrow)
YFP
Reds (quad)
Reds (oly MNG2)
Cy5 (quad narrow)
Cy5 (wide)
Fura-2
Cameleons 2
DIC

Stage heater and perfusion.

We have a homemade heater block that accommodates a standard 35mm dish. The heavy aluminum shell maintains
the dish at set point with constant circulation of water from the bath. This is normally set to 37°C or as high as 39°C to account
for thermal loss between the bath and stage. Use a digital thermometer to check the actual dish temperature, if critical. Heating
the dish will increase the rate of evaporation, a headache for long timelapse studies. If you are doing DIC imaging, you need to
avoid plastic since it degrades the DIC effect. One solution is to use a dish that is fitted with a coverslip as the bottom. You will
need to seal the top for long experiments and you may then need a cover also fitted with glass windows. Moisture will
condense over time, so a double pane design may work better. Another possibility is to perfuse the dish using a pump or
gravity-feed system. It is critical that you consult us on how to set this up so as to avoid flooding the system. If you have a
spill, let us know right away so that it can be cleaned up before damage to the microscope can occur. Do not think you can still
finish the experiment—it's finished! We need to track down leaks quickly since the salts will seep in and corrode/ruin
components. You can use a J-tube to maintain bath levels and to help safeguard against overflow leaks.
PDF Created with deskPDF PDF Writer - Trial :: http://www.docudesk.com
APPENDIX
exciter
380
360-370
436/20
480/40HQ
480/40HQ
500/20
530-550
530-550
640/20HQ
620/60HQ
340; 380
440/20
dichroic
emission
Quad
450/65
400
420LP
455
480/40
505
510LP
quad
535/30HQ
515
535/30HQ
Quad
630/60
570
590LP
Quad
682/22/75HQ
660
700/75HQ
400
510/40
455
535/35; 485/40
analyzer (polarizer)
31