Olympus IX70 Panduan Pengoperasian - Halaman 2

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Mikroskop Olympus IX70. Olympus IX70 32 halaman. Inverted fluorescent microscope
Juga untuk Olympus IX70: Petunjuk Pengoperasian (5 halaman)

Olympus IX70 Panduan Pengoperasian
1) fire the arc lamp under the microscope table
(make sure nothing else is on)
2) turn on the brightfield bulb power
switch on the microscope base (#2)
3) turn on the filter changer box (#3)
4) power up the computer (#4)
5) ONLY if you are using TIRF, fire the Ar laser
6) log on and start either MetaMorph or MetaFluor
7) write down your start time on the log sheet
The arc lamp must be on for at least 20-30 min prior to turning it off, you must wait that long before
turning it back on, too. Firing the arc greatly shortens the lamp life, so it's best to leave it running during
the day when the schedule shows other users are coming—check the schedule, the last person MUST shut
down the system.
If there is another user after you:
1) Clean any oil or water immersion objectives you used during your session.
2) Log off from your Windows session
3) If you used the TIRF laser (#5), turn that off; leave everything else (microscope and computer) on.
4) Write down the time on the log sheet.

If you are the last person scheduled:

1) Clean any oil or water immersion objectives you used during your session.
2) Log off your Windows session, then do another cntl-alt-del and choose the SHUTDOWN button.
3) Turn off the TIRF laser (#5) if used, the filter changer box (#3), brightfield power (#2), and the arc
lamp (#1).
4) Cover the scope with the plastic cover.
5) Write down the time on the log sheet (how else will the next person know if system is cool
enough to restart?)
PDF Created with deskPDF PDF Writer - Trial :: http://www.docudesk.com

Turning on the System

#1
always!
#2

Shutting down the System

#3
#4
#5 for TIRF
ONLY
2